No, the resolution of a spectrum depends only on the spectrometer’s resolution (slit width and diffraction grating used) and the sample’s alignment to the CL system. The beam current and accelerating voltage do not change the emitted spectral profile, but do influence the number of photons emitted so that it will have an impact on the spectrum’s signal-to-noise ratio.
In many insulator materials, including ceramics, rocks, minerals, and gems, CL can reveal the presence of many different trace elements. This detection capability is compatible with certain use cases and not recommended as a generic fingerprinting tool.
CL can map many of the rare earth elements at concentrations (sub-parts per million) that are far lower than are possible with EDS or WDS in the electron microprobe and detect transition metal cations (and their charge state). More recently, quantitative analysis of the titanium cation in quartz is useful for thermobarometry purposes.
SEM-based systems from Gatan do not use fibers. This allows the Monarc detector to preserve the angular information in the image projected from the collection mirror.